Proportional assist ventilation (PAV) versus neurally adjusted ventilator assist (NAVA): effect on oxygenation in infants with evolving or established bronchopulmonary dysplasia (2024)

  • Journal List
  • Springer Nature - PMC COVID-19 Collection
  • PMC7220976

As a library, NLM provides access to scientific literature. Inclusion in an NLM database does not imply endorsem*nt of, or agreement with, the contents by NLM or the National Institutes of Health.
Learn more: PMC Disclaimer | PMC Copyright Notice

Proportional assist ventilation (PAV) versus neurally adjusted ventilator assist (NAVA): effect on oxygenation in infants with evolving or established bronchopulmonary dysplasia (1)

Link to Publisher's site

Eur J Pediatr. 2020; 179(6): 901–908.

Published online 2020 Jan 25. doi:10.1007/s00431-020-03584-w

PMCID: PMC7220976

PMID: 31980954

Katie A. Hunt,1,2 Theodore Dassios,1,3 and Anne GreenoughProportional assist ventilation (PAV) versus neurally adjusted ventilator assist (NAVA): effect on oxygenation in infants with evolving or established bronchopulmonary dysplasia (2)1,2,4

Author information Article notes Copyright and License information PMC Disclaimer

Abstract

Both proportional assist ventilation (PAV) and neurally adjusted ventilatory assist (NAVA) provide pressure support synchronised throughout the respiratory cycle proportional to the patient’s respiratory demand. Our aim was to compare the effect of these two modes on oxygenation in infants with evolving or established bronchopulmonary dysplasia. Two-hour periods of PAV and NAVA were delivered in random order to 18 infants born less than 32weeks of gestation. Quasi oxygenation indices (“OI”) and alveolar-arterial (“A-a”) oxygen gradients at the end of each period on PAV, NAVA and baseline ventilation were calculated using capillary blood samples. The mean “OI” was not significantly different on PAV compared to NAVA (7.8 (standard deviation (SD) 3.2) versus 8.1 (SD 3.4), respectively, p = 0.70, but lower on both than on baseline ventilation (mean baseline “OI” 11.0 (SD 5.0)), p = 0.002, 0.004, respectively). The “A-a” oxygen gradient was higher on PAV and baseline ventilation than on NAVA (20.8 (SD 12.3) and 22.9 (SD 11.8) versus 18.5 (SD 10.8) kPa, p = 0.015, < 0.001, respectively).

Conclusion: Both NAVA and PAV improved oxygenation compared to conventional ventilation. There was no significant difference in the mean “OI” between the two modes, but the mean “A-a” gradient was better on NAVA.

What is Known:

Proportional assist ventilation (PAV) and neurally adjusted ventilatory assist (NAVA) can improve the oxygenation index (OI) in prematurely born infants.

• Both PAV and NAVA can provide support proportional to respiratory drive or demand throughout the respiratory cycle.

What is New:

• In infants with evolving or established BPD, using capillary blood samples, both PAV and NAVA compared to baseline ventilation resulted in improvement in the “OI”, but there was no significant difference in the “OI” on PAV compared to NAVA.

• The “alveolar-arterial” oxygen gradient was better on NAVA compared to PAV.

Open in a separate window

Keywords: Neurally adjusted ventilator assist, Proportional assist, Oxygenation index

Introduction

Mechanical ventilation can be live-saving in the neonatal period but is also associated with complications such as bronchopulmonary dysplasia (BPD) and chronic respiratory morbidity [1]. Modes of ventilation that allow synchrony both of the timing and the level of support to the infant’s respiratory effort have been developed. Both proportional assist ventilation (PAV) and neurally adjusted ventilator assist (NAVA) provide support synchronised throughout the respiratory cycle. PAV delivers support proportional to the infant’s respiratory effort. Inspiratory pressure can be delivered in proportion to the change in flow (resistive unloading) and the change in tidal volume (elastic unloading), and the clinician can adjust the amount of unloading used [2]. The support delivered by NAVA is proportional to the electrical activity of the diaphragm, which is reflective of the neural respiratory drive. NAVA uses a specialised nasogastric tube with an electrode array at the distal end which detects the electromyogram of the diaphragm (Edi). The Edi is the signal used to trigger the ventilator and determines the level of support. The delivered pressure throughout each inflation is in proportion to the Edi signal. The clinician can adjust the NAVA level to increase or decrease the amount of pressure delivered per microvolt of Edi detected [3, 4].

The results of short-term studies [515] suggest that both PAV and NAVA improved oxygenation and were associated with lower airway pressures compared to conventional or other triggered modes of ventilation in prematurely born infants with evolving BPD, otherwise known as chronic pulmonary insufficiency of prematurity (CPIP) [16]. There are no studies in that population comparing PAV and NAVA, and hence the aim of this study was to compare the effect of PAV and NAVA on the oxygenation index in infants with evolving or established BPD. It has been demonstrated in an in vitro study that PAV was associated with a relatively long trigger delay [17]. In studies in adults, longer trigger delays have been found with PAV compared to NAVA [18, 19]. Hence, we hypothesised that NAVA would be associated with a superior (i.e. lower) OI in a crossover study as more of the infant’s respiratory cycle would be supported by pressure support.

Methods

Infants were recruited from the tertiary neonatal unit at King’s College Hospital NHS Foundation Trust between June 2017 and July 2018. Infants were eligible for the study if they were born at a gestational age of less than 32weeks and remained ventilated at or beyond 1week after birth, that is, they had evolving or established BPD [20]. Those receiving neuromuscular blockade or with complex congenital cardiac abnormalities were excluded. The study was approved by the London-South East NHS Research Ethics Committee and the Health Research Authority and prospectively registered on clinicaltrials.gov with the identifier NCT02967549. Written informed parental consent was obtained.

Infants were routinely ventilated via shouldered Coles endotracheal tubes (Portex, Smith Medical, Hythe, UK) which have been shown to have minimal to no leak [21]. Prior to entry into the study, the infants were ventilated using SLE 5000 or 6000 ventilators (Specialised Laboratory Equipment, Croydon, UK) in assist control (A/C) or synchronised intermittent mandatory ventilation (SIMV) modes. Volume targeting was preferentially used, with target tidal volumes between 5 and 7ml/kg as per the Unit’s guidelines. Some clinicians, however, preferred to use pressure-limited ventilation. During the study, the infants were transferred to either the Servo-n ventilator (Maquet Critical Care, Solna, Sweden) or the Stephanie ventilator (Stephan GMBH, Gackenbach, Germany) depending on whether they had been randomised to receive first NAVA or PAV. The ventilator circuits were changed according to ventilator type. The randomisation sequence was generated with a random number generator and allocations concealed inside opaque, sealed and consecutively numbered envelopes. Infants received ventilation at their baseline settings on either ventilator, followed by 2h of either NAVA or PAV. At the end of the 2-hour period, a blood gas was taken, and the infant was transferred to the other study ventilator, following which the above sequence was repeated, that is, 1h on their baseline settings and 2h on the study mode. Infants were monitored throughout, and the heart rate, respiratory rate and oxygen saturations were recorded every 10min, and all desaturations less than 88% were documented.

NAVA was delivered by the Maquet Servo-n ventilator. A six French Edi catheter of appropriate length for the infant’s weight (49cm for 500–1500g, 50cm for 1000–2000g) was inserted oro- or nasogastrically prior to the commencement of the study and left in place until the study was completed. The Edi catheter was positioned as per the manufacturer’s guidance using the specialised tape measure in the packaging and correct position confirmed using the Edi catheter positioning guide function on the ventilator (Maquet Servo-n User Manual Version 1.1). The Edi was monitored using the Servo-n ventilator throughout the entire study, including on baseline settings, and whilst PAV was delivered via the Stephanie ventilator.

During NAVA, the NAVA level was set by observing the displayed pressure waveform on the ventilator during three to five breaths, whilst the infant was ventilated using their baseline settings and then adjusted so that the pressure on NAVA closely matched that delivered on baseline settings, aiming for a peak Edi of between 5 and 15μV. The apnoea time was set to 2s to avoid hypoventilation and the Edi trigger to 0.5μV as recommended by the manufacturer for neonatal use. The positive end-expiratory pressure (PEEP) was set at the same level as during baseline ventilation. Maximum PIP was initially limited to 5cmH2O above that set on baseline settings, as the Servo-n ventilator will not deliver pressures more than 5cmH2O below this limit. Ventilator parameters were manually collected with a single value that was expected to represent the existing stable baseline.

PAV was delivered using the Stephanie ventilator (Stephan GMBH, Gackenbach, Germany). Whilst ventilating on baseline settings, the infant’s compliance was recorded from the ventilator display every 10min. The mean of the six results was calculated to determine the degree of elastic unloading required to return the infant’s compliance to normal, that is, 2ml/cmH2O [22]. Initial elastic unloading was set at 75% of that value and increased to 100% after 10min if no abnormal waveforms were observed [17]. No resistive unloading was used as this can result in oscillation in the airway pressure waveform, which may inhibit diaphragm activity [23, 24]. The PEEP was set as the same level as during baseline settings.

The FiO2 was adjusted to maintain saturations between 92 and 96% as per the Unit’s guidelines. Expiratory tidal volume, FiO2, respiratory rate, MAP, PIP and PEEP were recorded from the ventilator display (no PIP or PEEP is displayed during PAV). Blood gas analysis was performed using capillary samples. Capillary samples were not formally arterialised. Blood gas samples were analysed immediately using the blood gas analyser on the neonatal unit (ABL90 Flex, Radiometer, Brønshøj, Denmark).

As capillary blood samples were used, we calculated quasi OIs and alveolar-arterial oxygen gradients which we have designated “OIs” and “A-a” gradient, respectively. The primary outcome, “OI”, was calculated as (MAP*FiO2*100)/(partial pressure of oxygen (pO2). Secondary outcomes included the “A-a” gradient calculated as (FiO2*(Patm – pH2O) – (PCO2/0.8)-pO2), where Patm = 101.33kPa and pH2O = 6.3kPa; the oxygen saturation to fraction of inspired oxygen ratio (SF ratio) calculated as SpO2/FiO2 and the PO2/FiO2 ratio. In addition, the peak and tonic Edi during each epoch were obtained from the Maquet ventilator.

BPD was diagnosed as oxygen dependency for at least 28days and as mild, moderate or severe depending on support required at 36weeks corrected gestational age as per the NIH consensus guidelines [25].

Sample size

A sample size of 18 infants was chosen as this would allow detection of one standardised difference in the results with 80% power at the 5% significance level. In a previous study [14], we reported a difference in the mean OI of infants supported by NAVA compared to ACV was greater than 3 with a standard deviation of the infants’ results being 2. We had also reported a difference of 3 in previous studies of PAV [6, 8]. Thus, the calculated sample size would be able to detect such a difference (i.e. >3) between the results on the two modes.

Statistical analysis

Results were compared from the end of each period of PAV, NAVA and conventional ventilation. Data were assessed for normality using a Shapiro-Wilk test. Normally distributed data were analysed using a one-way repeated measures ANOVA with Bonferroni correction. If the assumption of sphericity was violated, a Greenhouse-Geisser correction was applied. Non-normally distributed data were analysed using a Friedman test. Statistical analysis was carried out using IBM SPSS Statistics version 25 (IBM Corporation, Armonk, New York, USA.)

Results

Eighteen infants were studied (Table ​(Table1)1) (eight male, ten female). They were born at a median gestational age of 25.3 (range 23.6–30.3) weeks with median birthweight of 750 (range 454–950) grams. They were studied at a median of 20.5 (range 8–58) postnatal days. All of the infants were subsequently diagnosed with BPD; 13 had severe BPD, 3 moderate and 2 mild. Prior to the study, 15 infants were receiving volume-targeted ventilation, with target tidal volumes between 5 and 7ml/kg; the others were receiving pressure-limited ventilation. Nine infants were randomised to receive NAVA first (Fig.1). NAVA levels used ranged between 0.4 and 1.8.

Table 1

Baseline demographic details

N18
Sex (male)8
Gestational age (weeks)25.3 [23.6–30.3]
Birthweight (grams)750 [454–950]
Postnatal age at study (days)20.5 [8–58]
Weight at study (grams)865 [700–1800]

Open in a separate window

Results are presented as median [range]

Open in a separate window

Fig.1

Consort diagram of recruitment and study protocol

There was no significant difference in the mean “OI” at the end of 2h on PAV compared to on NAVA (mean 7.8 (standard deviation (SD) 3.2) versus 8.1 (SD 3.4), p = 0.7), but on both PAV and NAVA, the mean “OI” was significantly lower than on baseline ventilation (Table ​(Table2,2, Fig.2). The FiO2 was higher on PAV than on NAVA (mean 0.39 (SD 0.15) versus 0.37 (SD 0.13) p = 0.024) and the MAP lower on PAV than on NAVA (8.3 cmH2O (SD 1.1) versus 9.0 cmH2O (SD (0.9), p = 0.025)) (Table ​(Table2).2). If only those 13 infants who went on to develop severe BPD were included, the “OI” was significantly lower on NAVA or PAV than at baseline (8.4, 8.8 vs 12.8, p = <0.001, p = 0.023, respectively), but the “OI” was not significantly different on PAV and NAVA (p = 0.29).

Table 2

Comparison of ventilatory, blood gas and diaphragmatic electrical activity parameters between baseline ventilation, PAV and NAVA

BaselinePAVNAVApPairwise comparisons
“OI”11.0 [5.0]7.8 [3.2]8.1 [3.4]0.001

n-p 0.70

b-p 0.002

b-n 0.004

Mean airway pressure (cmH2O)9.5 [1.1]8.3 [1.1]9.0 [0.9]<0.001

n-p 0.025

b-p < 0.001

b-n 0.02

FiO20.40 [0.15]0.39 [0.15]0.37 [0.13]<0.001

n-p 0.024

b-p 0.28

b-n 0.003

“A-a” gradient (kPa)22.9 (11.8)20.8 (12.3)18.4 (10.8)<0.001

n-p 0.015

b-p 0.127

b-n < 0.001

SpO2/FiO2268 [96]268 [92]295 [102]0.002

n-p 0.04

b-p 1.0

b-n 0.04

PO2/FiO2104 (35)120 (40)126 (42)0.001

n-p 0.79

b-p 0.033

b-n 0.002

PCO2 [kPa]8.1 [1.7]8.9 [1.8]8.8 [1.7]0.026

n-p 1.0

b-p 0.007

b-n 0.16

PO2 [kPa]5.4 [0.76]5.66 [0.68]5.6 [1.0]0.69
Respiratory rate (breaths/min)61 [9]66 [8]56 [9]0.001

n-p 0.003

b-p 0.19

b-n 0.093

Expiratory tidal volume (ml/kg)6.2 [0.7]6.4 [1.0]6.5 [1.2]0.32
Peak inspiratory pressure (cmH2O)*17.8 [3.5]14.3 [3.0]b-n < 0.001
Positive end-expiratory pressure (cmH2O)*5.6 [4.7–7]6 [5–7]b-n 0.94
pH7.31 [0.67]7.29 [0.68]7.3 [0.66]0.053
Peak electrical activity of the diaphragm (microvolts)13.6 [5.8–40.4]13.1 [2.4–38.4]11.3 [5.2–28.5]0.33
Minimum electrical activity of the diaphragm (microvolts)2.2 [0.9–4.2]2.3 [0.7–4.2]2.4 [0.7–4.5]0.92

Open in a separate window

Results are presented as mean (standard deviation) or median [range]

n-p = comparison between NAVA and PAV; b-p = comparison between baseline and PAV

b-n = comparison between baseline and NAVA

*Data unavailable for two infants

Open in a separate window

Fig.2

“Oxygenation index” of individual infants shown by linked data points on baseline ventilation, PAV and NAVA

The “A-a” gradient was significantly higher on PAV than on NAVA (20.8 versus 18.4, p = 0.015.) It was also significantly lower on NAVA than at baseline (18.4 versus 22.9, p < 0.001), but there was no statistically significant difference between PAV and baseline (20.8 versus 22.9, p = 0.127) (Table ​(Table2).2). If the 13 infants who went on to develop severe BPD were analysed separately, both PAV and NAVA were associated with lower “A-a” gradients compared to baseline ventilation (23.2 (SD 13.7) and 20.9 (SD 11.4) versus 26.5 (SD 12.1) kPa, p = 0.031, 0.001, respectively), but there was no statistically significant difference between the results on PAV and NAVA (p = 0.18). The SpO2/FiO2 ratio was lower on PAV and baseline than on NAVA (p = 0.004) (Table ​(Table2).2). There was no significant difference in the PO2/FiO2 between NAVA and PAV, but both were higher than at baseline (Table ​(Table22).

There were no significant differences between the expiratory tidal volumes or the pCO2 levels on PAV compared to on NAVA (Table ​(Table2).2). There were no significant differences between base excess or HCO3 results on PAV and NAVA. The PCO2 was statistically significantly higher on PAV than at baseline (8.9 vs 8.1kPa, p = 0.016). The respiratory rate was significantly higher on PAV than on NAVA. The peak and minimum Edi were not significantly different between modes of ventilation (Table ​(Table22).

Discussion

We have demonstrated that there was no significant difference in mean “OI” between PAV and NAVA, but both modes resulted in a reduction in “OI” compared to baseline conventional ventilation. We used the “OI” as the primary outcome as this gives a summary of the effectiveness of each ventilator mode compared to baseline. The mean airway pressure, however, was lower and the FiO2 higher on PAV than on NAVA. As the “OI” is a composite of the MAP and FiO2, then those differences being in the opposite direction resulted in similar “OIs”. The lower MAP on PAV resulted in a higher FiO2 requirement to maintain the same level of oxygenation. The lower MAP on PAV might reflect lower levels of synchrony, a different shape to the airway pressure waveform and/or a longer trigger delay [1719]. The significantly higher “A-a” gradient (i.e. worse) on PAV compared to NAVA is consistent with the higher FiO2 requirement, as is the significantly lower SpO2/FiO2 (ie worse). The PO2/FiO2 was lower on PAV, but this did not reach statistical significance.

Both PAV and NAVA have been compared to conventional and other triggered modes of ventilation in small short-term studies in infants. PAV as compared to pressure-controlled conventional ventilation has been shown to reduce the OI in infants with respiratory distress syndrome [5] and to reduce the OI, peak and MAP and the work of breathing (as measured by PTPdi) in infants with evolving or established BPD [68]. Although there have been no randomised controlled trials of PAV in neonates, in adults, PAV versus pressure support ventilation resulted in a shorter duration of intensive care days [9]. NAVA has also been shown to have advantages over conventional ventilation in prematurely born infants. NAVA compared to pressure control ventilation and to pressure-limited synchronised intermittent mandatory ventilation (SIMV) in crossover studies, including those where infants had longer durations on each mode, resulted in lower peak inspiratory pressures (PIP) and a reduction in the work of breathing [1013]. In prematurely born infants with evolving or established BPD, NAVA compared to pressure control ventilation in a crossover study resulted in reduction in the OI, MAP, FiO2 and PIP [14]. There has been one randomised controlled trial of NAVA, which was compared to pressure control ventilation and enrolled infants born between 28 and 36 + 6weeks gestational age with acute respiratory distress. There was no significant difference in the primary outcome of the duration of ventilation (35 versus 26h, p = 0.21). There was also no difference in the incidence of BPD, but the study was not powered to detect this [15].

To our knowledge, there have been no previous studies in the neonatal population that have compared NAVA and PAV, but a study in adults compared NAVA and PAV to baseline pressure support ventilation [19]. The OI was not reported, but when targeting the same tidal volumes, the PIP and MAP were higher on NAVA than on PAV with comparable blood gases, although the PaCO2 was statistically significantly higher on PAV than on NAVA, as was also found in our study. The small increase in PaCO2 may be due to the way that NAVA and PAV switch to back up ventilation after an apnoea period or to the amount of unloading during PAV and the NAVA level set. If either were to be used for longer periods, then careful monitoring and adjustment of settings as required would be advised.

There are strengths and some limitations to our study. We calculated the “OI” and “A-a” gradients from capillary samples, but this was a crossover study, so each infant acted as their own control. Thus, whilst arterial samples may have been more precise, we suggest that use of capillary blood samples would not have substantially influenced our results. We did not use arterialisation of the capillary samples as there is controversy in the literature as to whether this improves the correlations with arterial results. Furthermore, capillary blood gases have been shown to correlate well with arterial gases if infants are stable and not cardiovascularly compromised [26], particularly at lower levels of arterial oxygenation as in our cohort [27]. Additionally, it has been shown that in infants with congenital diaphragmatic hernia, indices of oxygenation obtained from capillary blood gas samples and changes in these indices over time predict survival, thus suggesting that capillary oxygenation is a valid method of monitoring clinical status [28]. Nevertheless, for future studies, transcutaneous monitoring of carbon dioxide and oxygen levels could be considered. We did not assess the infants’ comfort levels, and this plus assessment of the degree of asynchrony would be useful in future studies.

The NAVA level was set as per the manufacturer’s guidelines. Using the manufacturer’s guidelines, the MAP level during NAVA was decreased compared to baseline setting suggesting that the level was appropriate. All of the infants in our study either had or were subsequently diagnosed with BPD, indicating that our inclusion criteria successfully identified those with evolving or established BPD. The wide range of postnatal ages in our study population may have influenced our results, but advantages of PAV and NAVA over conventional ventilation have been demonstrated in infants with a variety of postnatal ages in previous studies. Targeted tidal volumes of between 5 and 7mls/kg were used, as studies have shown chronically ventilated infants may require 7mls/kg [29, 30] and in the latter study (30) to reduce the work of breathing. It could be suggested that in babies with BPD 2-hour epochs may not be sufficient to demonstrate changes in oxygenation. Previous studies, however, have demonstrated changes in oxygenation using both modes for shorter epochs of 45–60min [5, 6, 14]. Indeed, we were able to demonstrate improved oxygenation over baseline settings for both PAV and NAVA using 2h epochs and differences in the two modes with regard to the A-a gradient. In one study [12], 12-h periods were used, but the order was not randomised. Although there were no significant differences in blood gases between NAVA and pressure-regulated volume control (PRVC), neural apnoeas were only seen during PRVC, and less fentanyl was administered during NAVA (12).

In conclusion, we have found no significant differences in the “OI” after 2h of support by PAV or NAVA but both significantly improved the OI compared to baseline ventilation. Other indices, as the ‘A-a’ gradient and SF ratio demonstrated that NAVA improved oxygenation over PAV, but the differences were relatively small.

Abbreviations

A-aAlveolar-arterial gradient
BPDBronchopulmonary dysplasia
CPIPChronic pulmonary insufficiency of prematurity
EdiElectromyogram of the diaphragm
FiO2Inspired oxygen concentration
MAPMean airway pressure
NAVANeurally adjusted ventilatory assist
OIOxygenation index
PAVProportional assist ventilation
PEEPPositive end-expiratory pressure
PIPPeak inspiratory pressure
PRVCPressure-regulated volume control
VTeExpiratory tidal volume

Contributors’ statement

AG designed the study, analysed the data and produced the final manuscript.

KH collected the data, analysed the data and produced the final manuscript.

TD analysed the data and produced the final manuscript.

Funding information

KAH was supported by the Charles Wolfson Charitable Trust and additionally by SLE. This research was supported by the National Institute for Health Research (NIHR) Biomedical Research Centre at Guy’s and St Thomas’ NHS Foundation Trust and King’s College London. The views expressed are those of the authors and not necessarily those of the NHS, the NIHR or the Department of Health. The Stephanie ventilator (PAV) and the Macquet Servo-n ventilator (NAVA) were loaned to us.

Compliance and ethical standards

All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.

Conflict of interest

Professor Greenough has held grants from various manufacturers (Abbot Laboratories, MedImmune) and ventilator manufacturers (SLE). Professor Greenough has received honoraria for giving lectures and advising various manufacturers (Abbot Laboratories, MedImmune) and ventilator manufacturers (SLE). Professor Greenough is currently receiving a non-conditional educational grant from SLE.

Informed consent

Parents gave informed written consent.

Footnotes

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Contributor Information

Katie A. Hunt, Email: ku.ca.lck@tnuh.a.eitaK.

Theodore Dassios, Email: u.ca.lck@soissad.erodoehT.

Anne Greenough, Email: ku.ca.lck@hguoneerg.ennA.

References

1. Fischer HS, Buhrer C. Avoiding endotracheal ventilation to prevent bronchopulmonary dysplasia: a meta-analysis. Pediatrics. 2013;132:e1351–e1360. doi:10.1542/peds.2013-1880. [PubMed] [CrossRef] [Google Scholar]

2. Schulze A. Respiratory mechanical unloading and proportional assist ventilation in infants. Acta Paediatr. 2007;91:19–22. doi:10.1111/j.1651-2227.2002.tb00155.x. [PubMed] [CrossRef] [Google Scholar]

3. Stein H, Firestone K. Application of neurally adjusted ventilatory assist in neonates. Semin Fetal Neonatal Med. 2014;19:60–69. doi:10.1016/j.siny.2013.09.005. [PubMed] [CrossRef] [Google Scholar]

4. Stein H, Firestone K, Rimensberger PC. Synchronized mechanical ventilation using electrical activity of the diaphragm in neonates. Clin Perinatol. 2012;39:525–542. doi:10.1016/j.clp.2012.06.004. [PubMed] [CrossRef] [Google Scholar]

5. Schulze A, Gerhardt T, Musante G, Schaller P, Claure N, Everett R, Gomez-Marin O, Bancalari E. Proportional assist ventilation in low birth weight infants with acute respiratory disease: a comparison to assist/control and conventional mechanical ventilation. J Pediatr. 1999;135:339–344. doi:10.1016/S0022-3476(99)70131-9. [PubMed] [CrossRef] [Google Scholar]

6. Bhat P, Patel DS, Hannam S, Rafferty GF, Peaco*ck JL, Milner AD, Greenough A. Crossover study of proportional assist versus assist control ventilation. Arch Dis Child Fetal Neonatal Ed. 2015;100:F35–F38. doi:10.1136/archdischild-2013-305817. [PubMed] [CrossRef] [Google Scholar]

7. Schulze A, Rieger-Fackeldey E, Gerhardt T, Claure N, Everett R, Bancalari E. Randomized crossover comparison of proportional assist ventilation and patient triggered ventilation in extremely low birth weight infants with evolving chronic lung disease. Neonatology. 2007;92:1–7. doi:10.1159/000098376. [PubMed] [CrossRef] [Google Scholar]

8. Shetty S, Bhat P, Hickey A, Peaco*ck JL, Milner AD, Greenough A. Proportional assist versus assist control ventilation in premature infants. Eur J Pediatr. 2016;175:57–61. doi:10.1007/s00431-015-2595-4. [PubMed] [CrossRef] [Google Scholar]

9. Bosma KJ, Read BA, Bahrgard Nikoo MJ, Jones PM, Priestap FA, Lewis JF. A pilot randomized trial comparing weaning from mechanical ventilation on pressure support versus proportional assist ventilation. Crit Care Med. 2016;44:1098–1108. doi:10.1097/CCM.0000000000001600. [PubMed] [CrossRef] [Google Scholar]

10. Chen Z, Luo F, Ma XL, Lin HJ, Shi LP, Du LZ. Application of neurally adjusted ventilatory assist in preterm infants with respiratory distress syndrome. Zhongguo Dang Dai Er Ke Za Zhi. 2013;15:709–712. [PubMed] [Google Scholar]

11. Lee J, Kim HS, Sohn JA, Lee JA, Choi CW, Kim EK, Kim BI, Choi JH. Randomized crossover study of neurally adjusted ventilatory assist in preterm infants. J Pediatr. 2012;161:808–813. doi:10.1016/j.jpeds.2012.04.040. [PubMed] [CrossRef] [Google Scholar]

12. Longhini F, Ferrero F, De Luca D, Cosi G, Alemani M, Colombo D, Cammarota G, Berni P, Conti G, Bona G, Della Corte F, Navalesi P. Neurally adjusted ventilatory assist in preterm neonates with acute respiratory failure. Neonatology. 2015;107:60–67. doi:10.1159/000367886. [PubMed] [CrossRef] [Google Scholar]

13. Stein H, Alosh H, Ethington P, White DB. Prospective crossover comparison between NAVA and pressure control ventilation in premature neonates less than 1500 grams. J Perinatol. 2013;33:452–456. doi:10.1038/jp.2012.136. [PubMed] [CrossRef] [Google Scholar]

14. Shetty S, Hunt K, Peaco*ck J, Ali K, Greenough A. Crossover study of assist control ventilation and neurally adjusted ventilatory assist. Eur J Pediatr. 2017;176:509–513. doi:10.1007/s00431-017-2866-3. [PubMed] [CrossRef] [Google Scholar]

15. Kallio M, Koskela U, Peltoniemi O, Kontiokari T, Pokka T, Suo-Palosaari M, Saarela T. Neurally adjusted ventilatory assist (NAVA) in preterm newborn infants with respiratory distress syndrome—a randomized controlled trial. Eur J Pediatr. 2016;175:1175–1183. doi:10.1007/s00431-016-2758-y. [PubMed] [CrossRef] [Google Scholar]

16. Steinhorn R, Davis JM, Gopel W, Jobe A, Abman S, Laughon M, Bancalari E, Aschner J, Ballard R, Greenough A, et al. Chronic pulmonary insufficiency of prematurity: developing optimal endpoints for drug development. J Pediatr. 2017;191:15–21. doi:10.1016/j.jpeds.2017.08.006. [PubMed] [CrossRef] [Google Scholar]

17. Patel DS, Rafferty GF, Hannam S, Lee S, Milner AD, Greenough A. In vitro assessment of proportional assist ventilation. Arch Dis Child Fetal Neonatal Ed. 2010;95:F331–F337. doi:10.1136/adc.2009.170787. [PubMed] [CrossRef] [Google Scholar]

18. Akoumianaki E, Prinianakis G, Kondili E, Malliotakis P, Georgopoulos D. Physiologic comparison of neurally adjusted ventilator assist, proportional assist and pressure support ventilation in critically ill patients. Respir Physiol Neurobiol. 2014;203:82–89. doi:10.1016/j.resp.2014.08.012. [PubMed] [CrossRef] [Google Scholar]

19. Schmidt M, Kindler F, Cecchini J, Poitou T, Morawiec E, Persichini R, Similowski T, Demoule A. Neurally adjusted ventilatory assist and proportional assist ventilation both improve patient-ventilator interaction. Crit Care. 2015;19:56. doi:10.1186/s13054-015-0763-6. [PMC free article] [PubMed] [CrossRef] [Google Scholar]

20. Hunt KA, Dassios T, Ali K, Greenough A. Prediction of bronchopulmonary dysplasia development. Arch Dis Child Fetal Neonatal Ed. 2018;103:F598–F599. doi:10.1136/archdischild-2018-315343. [PubMed] [CrossRef] [Google Scholar]

21. Hird M, Greenough A, Gamsu H. Gas trapping during high frequency positive pressure ventilation using conventional ventilators. Early Hum Dev. 1990;22:51–56. doi:10.1016/0378-3782(90)90025-E. [PubMed] [CrossRef] [Google Scholar]

22. Migdal M, Dreizzen E, Praud JP, Vial M, Dehan M, Chambille B, Gaultier C. Compliance of the total respiratory system in healthy preterm and full-term newborns. Pediatr Pulmonol. 1987;3:214–218. doi:10.1002/ppul.1950030404. [PubMed] [CrossRef] [Google Scholar]

23. Chowdhury O, Bhat P, Rafferty G, Hannam S, Milner AD, Greenough A. In vitro assessment of the effect of proportional assist ventilation on the work of breathing. Eur J Pediatr. 2016;175:639–643. doi:10.1007/s00431-015-2673-7. [PubMed] [CrossRef] [Google Scholar]

24. Schulze A, Rich W, Schellenberg L, Schaller P, Heldt GP. Effects of different gain settings during assisted mechanical ventilation using respiratory unloading in rabbits. Pediatr Res. 1998;44:132–138. doi:10.1203/00006450-199807000-00021. [PubMed] [CrossRef] [Google Scholar]

25. Jobe AH, Bancalari E. NICHD / NHLBI / ORD workshop summary. Am J Respir Crit Care Med. 2001;163:1723–1729. doi:10.1164/ajrccm.163.7.2011060. [PubMed] [CrossRef] [Google Scholar]

26. Yildizdaş D, Yapicioğlu H, Yilmaz HL, Sertdemir Y. Correlation of simultaneously obtained capillary, venous, and arterial blood gases of patients in a paediatric intensive care unit. Arch Dis Child. 2004;89:176–180. doi:10.1136/adc.2002.016261. [PMC free article] [PubMed] [CrossRef] [Google Scholar]

27. McLain BI, Evans J, Dear PR. Comparison of capillary and arterial blood gas measurements in neonates. Arch Dis Child. 1988;63:743–747. doi:10.1136/adc.63.7_Spec_No.743. [PMC free article] [PubMed] [CrossRef] [Google Scholar]

28. Grizelj R, Bojanic K, Pritisanac E, Luetic T, Vukovic J, Weingarten TN, Schroeder DR, Sprung J. Survival prediction of high-risk outborn neonates with congenital diaphragmatic hernia from capillary blood gases. BMC Pediatr. 2016;16:114. doi:10.1186/s12887-016-0658-y. [PMC free article] [PubMed] [CrossRef] [Google Scholar]

29. Kezsler M, Nassabeh-Montazami S, Abubakar K. Evolution of tidal volume requirement during the first 3 weeks of life in infants <800 g ventilated with volume guarantee. Arch Dis Child Fetal Neonatal Ed. 2009;94:F279–F282. doi:10.1136/adc.2008.147157. [PubMed] [CrossRef] [Google Scholar]

30. Hunt KA, Dassios T, Ali K, Greenough A. Volume targeting levels and work of breathing in infants with evolving or established bronchopulmonary dysplasia. Arch Dis Child Fetal Neonatal Ed. 2019;104:F46–F49. doi:10.1136/archdischild-2017-314308. [PubMed] [CrossRef] [Google Scholar]

Articles from European Journal of Pediatrics are provided here courtesy of Nature Publishing Group

Proportional assist ventilation (PAV) versus neurally adjusted ventilator assist (NAVA): effect on oxygenation in infants with evolving or established bronchopulmonary dysplasia (2024)
Top Articles
Latest Posts
Article information

Author: Neely Ledner

Last Updated:

Views: 5943

Rating: 4.1 / 5 (42 voted)

Reviews: 81% of readers found this page helpful

Author information

Name: Neely Ledner

Birthday: 1998-06-09

Address: 443 Barrows Terrace, New Jodyberg, CO 57462-5329

Phone: +2433516856029

Job: Central Legal Facilitator

Hobby: Backpacking, Jogging, Magic, Driving, Macrame, Embroidery, Foraging

Introduction: My name is Neely Ledner, I am a bright, determined, beautiful, adventurous, adventurous, spotless, calm person who loves writing and wants to share my knowledge and understanding with you.